Monte Carlo dosimetry study of novel rotating MRI-compatible shielded tandems for intensity modulated cervix brachytherapy

Marc Morcos & Shirin A. Enger · 2020-03-09

Intensity modulated brachytherapy (IMBT) with rotating metal shields enables dose modulation that can better conform to the tumor while reducing OAR doses. In this work, we investigate novel rotating shields, compatible with MRI-compatible tandems used for cervix brachytherapy. Three unique shields were evaluated using the traditional Three different IMBT shields were modeled and simulated in RapidBrachyMCTPS. Each tungsten shield was designed to fit inside a 6 mm-wide MRI-compatible tandem. The active core of the source was replaced with TFs are favorable for IMBT and ranged between 12.9% and 32.2% for This study has quantitatively assessed the influence of various rotating cervical cancer-specific IMBT tandem shields on dosimetry. The dynamic single-channel shields and narrow beam-widths in the polar and azimuthal direction give rise to highly anisotropic distributions. Intermediate-to-high energy radionuclides,